Anode heel effect application with step wedge against effect of signal to noise ratio in computed radiography

https://doi.org/10.29332/ijhs.v4n3.467

Authors

  • Ni Nyoman Ratini Department of Physics, Faculty of Mathematics and Natural Sciences, Udayana University, Denpasar, Indonesia
  • I Made Yuliara Department of Physics, Faculty of Mathematics and Natural Sciences, Udayana University, Denpasar, Indonesia
  • Windaryoto Windaryoto Department of Physics, Faculty of Mathematics and Natural Sciences, Udayana University, Denpasar, Indonesia

Keywords:

AHE, computed radiography, image, SNR, step wedge

Abstract

It was researched on the application of Anoda Heel Effect (AHE) with a step wedge on the effect of Signal To Noise Ratio (SNR) on Computed Radiography (CR) has been carried out. This research was conducted on a 21 step wedge with two treatments, namely the application of AHE and without the application of AHE. This measurement is repeated three times on radiographs to obtain a total image of six images (without the application of AHE as many as three images and with the application of AHE as many as three images). The results of taking radiographs using AHE and without AHE were measured using the RadiAnt Dicom Viewer program. The SNR value on the step wedge image without the AHE application has an average of 26.89. The SNR value on the step wedge image using AHE is 60.54. The results of the correlation test (Pearson correlation test) on the SNR showed that there was a significant and very strong effect of the application of AHE on the step wedge on the SNR in CR (p-value <0.001 and the R-value ranging from 0.600 to 0.799).

Downloads

Download data is not yet available.

References

Carlton, R. R., & Adler, A. M. (2012). Principles of Radiographic Imaging (Book Only). Cengage Learning.

Cowen, A. R., Davies, A. G., & Kengyelics, S. M. (2007). Advances in computed radiography systems and their physical imaging characteristics. Clinical radiology, 62(12), 1132-1141. https://doi.org/10.1016/j.crad.2007.07.009

Dahlan, M. S. (2011). Statistik untuk kedokteran dan kesehatan. Penerbit Salemba.

Desai, N., Singh, A., & Valentino, D. J. (2010). Practical evaluation of image quality in computed radiographic (CR) imaging systems. In Medical Imaging 2010: Physics of Medical Imaging (Vol. 7622, p. 76224Q). International Society for Optics and Photonics.

Fauber, T. L. (2016). Radiographic Imaging and Exposure-E-Book. Elsevier Health Sciences.

Gede, P. S. L., Ratini, N. N., & Iffah, M. (2020). Pengaruh Variasi Tegangan Tabung Sinar-X terhadap Signal to Noise Ratio (SNR) dengan Penerapan Anode Heel Effect menggunakan Stepwedge. BULETIN FISIKA, 22(1), 20-28.

Gibson, D. J., & Davidson, R. A. (2012). Exposure creep in computed radiography: a longitudinal study. Academic radiology, 19(4), 458-462. https://doi.org/10.1016/j.acra.2011.12.003

Kenneth, L. B. (2014). Textbook of radiographic positioning and related anatomy. Elsevier mosby.

Korner, M., Weber, C. H., Wirth, S., Pfeifer, K. J., Reiser, M. F., & Treitl, M. (2007). Advances in digital radiography: physical principles and system overview. Radiographics, 27(3), 675-686.

Larsson, E. G., Erdogmus, D., Yan, R., Principe, J. C., & Fitzsimmons, J. R. (2003). SNR-optimality of sum-of-squares reconstruction for phased-array magnetic resonance imaging. Journal of Magnetic Resonance, 163(1), 121-123. https://doi.org/10.1016/S1090-7807(03)00132-0

Louk, A. C., & Suparta, G. B. (2014). Pengukuran Kualitas Sistem Pencitraan Radiografi Digital Sinar-X. BIMIPA, 24(2), 149-166.

Ma, W. K., Hogg, P., Tootell, A., Manning, D., Thomas, N., Kane, T., ... & Kitching, J. (2013). Anthropomorphic chest phantom imaging–the potential for dose creep in computed radiography. Radiography, 19(3), 207-211. https://doi.org/10.1016/j.radi.2013.04.002

Maintz, J. A., & Viergever, M. A. (1998). A survey of medical image registration. Medical image analysis, 2(1), 1-36. https://doi.org/10.1016/S1361-8415(01)80026-8

McInerney, T., & Terzopoulos, D. (1996). Deformable models in medical image analysis: a survey. Medical image analysis, 1(2), 91-108. https://doi.org/10.1016/S1361-8415(96)80007-7

Medixant. (2016). RadiAnt DICOM Viewer. Version 3.0.2

Ratini, N. N., Yuliara, I. M., & Trisnawati, N. L. P. (2019). Radiation dosage distribution in area room CT scan multi slice 64 to dose limit value. International journal of health sciences, 3(3), 25-32.

Sudin, A., Widyandari, H., & Muhlisin, Z. (2015). Studi Pengaruh Ukuran Pixel Imaging Plate Terhadap Kualitas Citra Radiograf. Youngster Physics Journal, 4(3), 225-230.

Published

15-11-2020

How to Cite

Ratini, N. N. ., Yuliara, I. M. ., & Windaryoto, W. (2020). Anode heel effect application with step wedge against effect of signal to noise ratio in computed radiography. International Journal of Health Sciences, 4(3), 75–82. https://doi.org/10.29332/ijhs.v4n3.467

Issue

Section

Peer Review Articles